| Abstract |
|---|
|
|
|---|
Method and Results— Between 2005 and 2008, we have enrolled 136 consecutive patients with percutaneous aortic valve replacement using the CoreValve prosthesis. In this prospective nonrandomized, single-center trial, we analyzed procedural outcome, complications and clinical status up to 1 year. First, second, and third generation of the CoreValve prosthesis were implanted in 10, 24, and 102 consecutive high-risk patients (logistic EuroScore: 23.1±15.0%) with severe symptomatic aortic valve stenosis. Mean transvalvular pressure gradient was 41.5±16.7 mm Hg. The procedural success rate increased from generation 1/2 to 3 from 70.0%/70.8% to 91.2% (P=0.003). The 30-day combined rate of death/stroke/myocardial infarction was 40.0%/20.8%/14.7% (P=0.11) for generation 1, 2, and 3, with no procedural death in generation 3. Pressure gradients improved significantly with a final mean gradient of 8.1±3.8 mm Hg. Overall functional status assessed by New York Heart Association class improved from 3.3±0.5 (pre) to 1.7±0.7 (post) (P<0.001) and remained stable in the follow-up.
Conclusion— In experienced hands, percutaneous aortic valve replacement with the CoreValve system for selected patients with severe aortic valve stenosis has a high acute success rate associated with a low periprocedural mortality/stroke rate as well as remarkable clinical and hemodynamic improvements, which persist over time. Additional studies are now required to confirm these findings, particularly head-to-head comparisons with surgical valve replacement in different risk populations.
Key Words: aortic valve disease aortic stenosis valve prosthesis percutaneous approach
| Introduction |
|---|
|
|
|---|
In this article, we report the procedural and 12 months follow-up experience with 3 generations of the self-expanding CoreValve Revalving aortic valve prosthesis consecutively implanted in a single high-volume center. Of particular interest is the outcome of the 18F generation device that is presently the commercially available product and the comparison of this device with the previous generations. Multicenter results for generations 1 and 2 have already been reported by our group.14,16
| Methods |
|---|
|
|
|---|
Inclusion criteria were the following: (I) severe native aortic valve stenosis with an valve area <1 cm2, with or without aortic valve regurgitation and (1) age
80 years or a logistic EuroSCORE
20% for the 25/21F group and age
75 years or logistic EuroSCORE
15% for 18F group, respectively; or (2) age
65 years with EuroSCORE
15% and at least one of the following complicating criteria: cirrhosis of liver, pulmonary insufficiency (FEV1<1 L), previous cardiac surgery, pulmonary hypertension >60 mm Hg, porcelain aorta, recurrent pulmonary embolus, right ventricular insufficiency. (II) Echocardiographic aortic valve annulus diameter
20 mm and
27 mm; (III) diameter of the ascending aorta
45 mm at the sino-tubular junction.
Preinterventional patient screening included transthoracic as well as transesophageal echocardiography to confirm diagnosis, duplex ultrasonography to assess the access site, multislice computer tomography to assess aortic and aortic valve dimensions and morphology (grade and distribution of calcifications, annulus dimension in a multiplanar reconstruction, measuring from hinge point to hinge point) as well as the access, and invasive cardiac evaluation with coronary angiogram, supra-aortic angiogram and left ventriculography. The baseline operative risk of the patients was estimated by the logistic EuroSCORE as well as the Society of Thoracic Surgeons (STS) score. The patient was considered high risk if the inclusion criteria were met as confirmed by an independent senior cardiologist and senior cardiac surgeon.
Aim of the Study
The purpose of the clinical trial reported here was to demonstrate the progress among the various CoreValve Revalving device generations and to evaluate the current feasibility, safety, and efficacy status up to 12 months postimplantation, particularly of the third generation 18F CoreValve Revalving prosthesis compared with device generations 1 (25F) and 2 (21F) based on the data from an experienced high-volume center.
Device Description
Design characteristics of the self-expanding CoreValve Revalving prosthesis as well as the procedural characteristics have been described previously in detail.13,15 Briefly, the CoreValve prosthesis consists of a trileaflet bioprosthetic porcine pericardial tissue valve, which is mounted and sutured in a self-expanding nitinol stent frame (Figure 1). The lower portion of the prosthesis has high radial force to expand and exclude the calcified leaflets and to avoid recoil. The middle portion is constrained to avoid the coronary arteries, whereas the upper portion is flared to center and fix the stent frame firmly in the ascending aorta and to provide longitudinal stability and coaxial positioning. The sizes of 3 subsequently developed delivery systems have been gradually reduced from 25F to 18F over time to facilitate vascular access and deployment of the device. With generation 3 (18F), there are 2 different device sizes available for different annulus dimensions: the 26-mm prosthesis for aortic valve annulus sizes from 20 to 24 mm and the 29 mm prosthesis for aortic valve annulus sizes from 24 to 27 mm.
|
After predilation, the prosthesis was deployed and implanted retrogradely within the aortic annulus over a stiff guide wire (Figure 2). Postdilatation of the CoreValve prosthesis was performed at the discretion of the operator depending on the perceived proper placement of the device angiographically and the aortic regurgitation grade. Clinical and echocardiographic follow-up was scheduled postprocedurally, at hospital discharge, 30 days, 6 and 12 months after device implantation. Routine postprocedural measurement of creatine kinase/creatine kinase-myocardial band was performed to detect myocardial ischemia. All patients provided written, informed consent before the procedure.
|
Statistical Analysis
Categorical variables are presented as frequencies and were compared by
2 test or Fisher exact test. Continuous variables are presented as mean±standard deviation. A 2-tailed unpaired Student t test for comparison of continuous data between groups and a paired Student t test for intragroup comparison was used. Ordinal variables were compared with the Mann-Whitney U test. A probability value of <0.05 was considered statistically significant. Analyses were conducted with SPSS version 13.0 (SPSS, Inc, Chicago, Ill). The authors had full access to and take full responsibility for the integrity of the data. All authors have read and agreed to the manuscript as written.
| Results |
|---|
|
|
|---|
|
|
|
Follow-Up Clinical Results
Follow-up clinical results are listed in Table 4. The overall mortality and major adverse cardiovascular and cerebral events (MACCE) rate at 30 days were 40%/8.3%/10.8% and 40.0%/20.8%/14.7% in the generation 1/2/3 groups, respectively. Overall functional status assessed by New York Heart Association class improved from 3.3±0.5 preprocedure to 1.7±0.7 postvalve implantation (P<0.001) without differences between the groups. Over a period of 12 months, the functional status remained stable in all 3 study groups. Beyond the 30-day follow-up, additional 8 deaths occurred, 3 in the 21F group and 5 in the 18F group. None of these were related to device dysfunction. There were no other late major adverse events reported in all 3 study groups (Figures 3 and 4
).
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Feasibility
Today, PAVR using the 18F CoreValve prosthesis is feasible and reliable in experienced hands with a high acute device success rate of about 97%. This is certainly related to an improved screening process on the one hand, which is crucial to identify suitable valvular and vessel morphologies. On the other hand, this reflects better device properties due to a reduced profile size as well as improved operator skills with growing knowledge on how to use the technique, where to deploy the device, and which landmarks are helpful to identify the correct landing zone (eg, calcifications of noncoronary cusp).
Multiple studies are currently ongoing to identify and scientifically validate screening tools to identify morphologically suitable patients. Multislice computer tomography assessment of the ventricle, aortic root, annulus, and aorta to femoral arteries is presently the most valuable method for screening patients. Factors that currently make the patients poor candidates for percutaneous implantation of the CoreValve prosthesis are femoral and iliac arteries <6 mm in diameter, severe kinking, and calcifications of the arterial course through which the valve has to pass, ventricles that are too horizontal when compared with the aortic root and annulus sizes that are too small or large. Calcification degrees and patterns of the aortic valve itself and their influence on success rates and outcome are currently subject for multiple ongoing studies.
The procedural learning curve over the past 3 years included the development of specific deployment techniques as well as various interventional bail-out strategies to deal with suboptimal initial PAVR results. The "step-deployment" technique, ie, temporarily halting device deployment after releasing the distal two third of the device (Figure 2) to ensure correct placement as opposed to rapid release of the entire prosthesis, allows normal blood flow as the prosthetic valve is already functioning while still having the option to pull the prosthesis either slightly for fine adjustments, or even completely to retract and remove the device. Postdilation using larger balloons can be used to increase the expansion of the prosthesis in case of paravalvular leakages or restriction of the outward expansion in the lower part of the prosthesis due to high radial inward forces from the native valve. Implantation of a second prosthesis within the first in case of suboptimal implantation of the first prosthesis too low within the native valve area can be performed and was safely done in 3 patients of the current 18F series with good hemodynamic results as assessed by angiogram and echocardiography. If the prosthesis is implanted slightly too deep toward the ventricular site, gentle pulling using a standard snare is another option for fine adjustments (used in 6 cases of the 18F series).
All these techniques as well as the ability to identify the right indication of using any of the above obviously require experienced operators and adequate logistics. Therefore, a rigorous training program is in place for new and existing sites to maintain these standards.
Safety
PAVR has been introduced to offer a safe treatment option for candidates in whom surgical aortic valve replacement is considered not to be safe, balancing perioperative operative risk versus the natural course of the disease/medical treatment. Therefore, mortality is the key safety parameter in all present PAVR studies. In our series, the periprocedural mortality decreased from 10% and 8% for the 25 F and 21F groups, which is in line with the previous publications on generations 1 and 2 CoreValve implantations14,16 to 0% for the 18F group, with a 30 day mortality of 9.8%. Given a highly comorbid population with a logistic EuroScore around 20% in all 3 groups, this result for procedural mortality is certainly remarkable and promising. The accuracy of risk predictor scores such as the popular EuroScore is certainly controversial, probably overestimating risk given recent advances in modern cardiovascular surgery. However, 0% periprocedural mortality as well as a one-digit 30-day mortality in this high-risk population indicates that PAVR is a safe and beneficial treatment modality for this patient population.
Reasons for death in the 30-day follow-up period in the 18F series were mainly related to comorbidities with only 2 events directly procedure-related caused by periprocedural tamponades. Tamponade due to myocardial perforation with the guide wire or a pacemaker lead, observed in this study and reported by other groups as well, is certainly a dangerous but fairly rare event the operator should be aware of. To avoid wire perforations, meticulous wire tip control in the ventricle is mandatory. During the course of the presently reported PAVR series, a manually shaped pigtail like wire tip has been identified to be less traumatic. This preshaping is now mandatorily implemented in the implantation process herewith successfully reducing the risk for myocardial perforations and subsequent tamponades.
Periprocedural strokes, initially observed in about 10% of patients with the first generations in our series and previously published series, is now around 3% with the 18F device, including 2% minor events. The successful reduction of these adverse events is potentially related to the profile reduction from 25F to 18F, which markedly facilitated the passage of the aortic arch as well as the insertability within the stenosed native valve.
The need for pacemaker implantations for treatment of AV conduction disturbances or postprocedural bradycardias, both procedure and medication related, increased from generations 1 to 3 in our center. This is certainly in parts related to our current local less restrictive implantation policy in case of any signs of imminent conduction problems. For instance, several patients received a permanent pacemaker already on the day of the PAVR procedure or the day after if there was any kind of documentation of a high-degree AV block or bradycardia related to slow atrial fibrillation in the first 2 to 6 hours after PAVR. In contrast, a more restrictive policy with a 48-hour period with a temporary pacemaker in place to give time for recovery has been followed in the beginning and is practice in other sites with comparable safety rates but lower rates of permanent pacemaker implantations. Reason for the policy change in our center was the intention to avoid complications more typically related to temporary pacing, such as fatal risk of lead dislocation in pacemaker-dependent patients.
Development of AV blocks after aortic valve interventions of any kind, either surgical replacement or percutaneous balloon valvuloplasty or percutaneous aortic valve implantations is a known risk given the proximity of atrio-ventricular conduction bundles and the aortic annulus/left ventricular outflow tract. Subtracting the cases with implantation of pacemakers due to nonprocedure-related bradycardias (medication related), the risk of directly PAVR-related pacemaker implants still appears slightly higher compared with surgical series with a reported pacemaker rate of 3.2% to 8.5%.17–19 Whether one PAVR method is more likely to cause atrio-ventricular conduction disturbances than the other is presently unclear. The fact is that this is usually a benign and well-controlled event given the current practice of mandatory periprocedural temporary pacemaker backup for 48 hours in all PAVR patients.
If the patient is event-free up to 30-day follow-up, the short and midterm adverse event rate up to 12 months is promising given the high-risk patient characteristics with an average age of 82 years in association with multiple comorbidities. There were no late adverse device-related events in this period in our trial.
Efficacy
Once the CoreValve Revalving prosthesis is successfully implanted, the hemodynamic status improves immediately, and this applies to all 3 device generations. Any preprocedural gradient can be reliably reduced to a typical mean pressure gradient of around 8 to 10 mm Hg with the device in place. Postprocedural mild aortic regurgitation is common, probably related to incomplete coverage of native valve commissures particularly in cases with compact focal calcifications. However, there were only 2 cases with a higher grade aortic regurgitation (3+) after device placement at the end of the procedure observed in our 18F series, which improved over time with medical or interventional treatment. Strategies to deal periprocedurally with regurgitations are described above including postdilations using large balloons or snare utilization to adjust device placement. Implantation of a second prosthesis within a suboptimally implanted prosthesis is another option to improve the outcome but should be considered only a bail-out concept not just due to the technical challenges but also due to effects such as unpredictable coronary access given 2 overlying nitinol net frames.
At follow-up, initially achieved benefits persisted over time. There were no events of late acquired significant paravalvular or valvular regurgitations or late increase of valvular gradients in any of the 3 study groups.
Overall, the comparison of the CoreValve generations presented here demonstrates that there was a learning curve for both engineers and clinicians. Comparing the first and second generation experience with the current device, which is commercially available, a remarkable reduction of periprocedural complications is detectable resulting in a higher procedural success rate and an improved midterm outcome. Given the smaller catheter size of 18F, there is no need for hemodynamic support, general anesthesia, or surgical support. The implantation is now solely guided by fluoroscopy with occasional contrast injections without additional visualization measures such as transesophageal echocardiography. This makes this procedure very tolerable for the patient, and reduces the procedure time to <45 minutes in experienced hands.
Study Limitations
The present study is a nonrandomized, single-center experience. Future studies are planned with conventional surgical as well as medical treatment arms as control. Follow-up of the patients is presently ongoing, so there was incomplete availability of midterm clinical and echocardiographic follow-up results particularly in the 18F group. Finally, the results to date also only apply to a high risk patient population as enrolled in this study. There are presently no data for patients with a standard low surgical risk profile.
| Conclusion |
|---|
|
|
|---|
| Acknowledgments |
|---|
The conduction of the study was supported by CoreValve, Inc.
Disclosures
Eberhard Grube is consultant to CoreValve. Ulrich Gerckens is clinical proctor for CoreValve. The other authors have nothing to disclose in relation to this article.
| References |
|---|
|
|
|---|
2. Varadarajan P, Kapoor N, Bansal R, Pai R. Clinical profile and natural history of 453 nonsurgically managed patients with severe aortic stenosis. Ann Thorac Surg. 2006; 82: 2111–2115.
3. Bramstedt KA. Aortic valve replacement in the elderly: frequently indicated yet frequently denied. Gerontology. 2003; 49: 46–49.[CrossRef][Medline]
4. Otto CM, Mickel MC, Kennedy JW, Alderman EL, Bashore TM, Block PC, Brinker JA, Diver D, Ferguson J, Holmes DR Jr. Three-year outcome after balloon aortic valvuloplasty: insights into prognosis of valvular aortic stenosis. Circulation. 1994; 89: 642–650.
5. Lieberman EB, Bashore TM, Hermiller JB, Wilson JS, Pieper KS, Keeler GP, Pierce CH, Kisslo KB, Harrison JK, Davidson CJ. Balloon aortic valvuloplasty in adults: failure of procedure to improve long-term survival. J Am Coll Cardiol. 1995; 26: 1522–1528.[Abstract]
6. Sack S, Kahlert P, Khandanpour S, Naber C, Philipp S, Möhlenkamp S, Sievers B, Kälsch H, Erbel R. Revival of an old method with new techniques: balloon aortic valvuloplasty of the calcified aortic stenosis in the elderly. Clin Res Cardiol. 2008; 97: 288–297.[CrossRef][Medline]
7. Sochman J, Peregrin JH, Pavcnik D, Timmermans H, Rosch J. Percutaneous transcatheter aortic disc valve prosthesis implantation: a feasibility study. Cardiovasc Intervent Radiol. 2000; 23: 384–388.[CrossRef][Medline]
8. Boudjemline Y, Bonhoeffer P. Steps toward percutaneous aortic valve replacement. Circulation. 2002; 105: 775–778.
9. Lutter G, Kuklinski D, Berg G, Von Samson P, Martin J, Handke M, Uhrmeister P, Beyersdorf F. Percutaneous aortic valve replacement: an experimental study. J Thorac Cardiovasc Surg. 2002; 123: 768–776.
10. Cribier A, Eltchaninoff H, Bash A, Borenstein N, Tron C, Bauer F, Derumeaux G, Anselme F, Laborde F, Leon MB. Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description. Circulation. 2002; 106: 3006–3008.
11. Webb JG, Munt B, Makkar RR, Naqvi TZ, Dang N. Percutaneous stent-mounted valve for treatment of aortic or pulmonary valve disease. Catheter Cardiovasc Interv. 2004; 63: 89–93.[CrossRef][Medline]
12. Cribier A, Eltchaninoff H, Tron C, Bauer F, Agatiello C, Sebagh L, Bash A, Nusimovici D, Litzler PY, Bessou JP, Leon MR. Early experience with percutaneous transcatheter implantation of heart valve prosthesis for the treatment of end-stage inoperable patients with calcific aortic stenosis. J Am Coll Cardiol. 2004; 43: 698–703.
13. Grube E, Laborde JC, Zickmann B, et al. First report on a human percutaneous transluminal implantation of a self-expanding valve prosthesis for interventional treatment of aortic valve stenosis. Catheter Cardiovasc Interv. 2005; 66: 465–469.[CrossRef][Medline]
14. Grube E, Laborde JC, Gerckens U, et al. Percutaneous implantation of the CoreValve self-expanding valve prosthesis in high-risk patients with aortic valve disease: the Siegburg first-in-man study. Circulation. 2006; 114: 1616–1624.
15. Webb JG, Chandavimol M, Thompson CR, Ricci DR, Carere RG, Munt BI, Buller CE, Pasupati S, Lichtenstein S. Percutaneous aortic valve implantation retrograde from the femoral artery. Circulation. 2006; 113: 842–850.
16. Grube E, Schuler G, Buellesfeld L, Gerckens U, Linke A, Wenaweser P, Sauren B, Mohr FW, Walther T, Zickmann B, Iversen S, Felderhoff S, Cartier R, Bonan R. Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second- and current third-generation self-expanding corevalve prosthesis: device success and 30-day clinical outcome. J Am Coll Cardiol. 2007; 50: 69–76.
17. Limongelli G, Ducceschi V, D'Andrea A, Renzulli A, Sarubbi B, De Feo M, Cerasuolo F, Calabrò R, Cotrufo M. Risk factors for pacemaker implantation following aortic valve replacement: a single centre experience. Heart. 2003; 89: 901–904.
18. Erdogan HB, Kayalar N, Ardal H, Omeroglu SN, Kirali K, Guler M, Akinci E, Yakut C. Risk factors for requirement of permanent pacemaker implantation after aortic valve replacement. J Card Surg. 2006; 21: 211–215.[CrossRef][Medline]
19. Dawkins S, Hobson AR, Kalra PR, Tang AT, Monro JL, Dawkins KD. Permanent pacemaker implantation after isolated aortic valve replacement: incidence, indications, and predictors. Ann Thorac Surg. 2008; 85: 108–112.
Related Article
Circ Cardiovasc Interv 2008 1: 159-160.
This article has been cited by other articles:
![]() |
L. Buellesfeld, P. Wenaweser, U. Gerckens, R. Mueller, B. Sauren, G. Latsios, B. Zickmann, G. Hellige, S. Windecker, and E. Grube Transcatheter aortic valve implantation: predictors of procedural success--the Siegburg-Bern experience Eur. Heart J., December 27, 2009; (2009) ehp570v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Bruschi, F. De Marco, J. Oreglia, P. Colombo, P. Fratto, F. Lullo, R. Paino, M. Frigerio, L. Martinelli, and S. Klugmann Percutaneous Implantation of CoreValve Aortic Prostheses in Patients With a Mechanical Mitral Valve. Ann. Thorac. Surg., November 1, 2009; 88(5): e50 - e52. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Bleiziffer, H. Ruge, D. Mazzitelli, A. Hutter, A. Opitz, R. Bauernschmitt, and R. Lange Survival after transapical and transfemoral aortic valve implantation: Talking about two different patient populations J. Thorac. Cardiovasc. Surg., November 1, 2009; 138(5): 1073 - 1080. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. F Tops, V. Delgado, F. van der Kley, and J. J Bax Percutaneous aortic valve therapy: clinical experience and the role of multi-modality imaging Heart, September 15, 2009; 95(18): 1538 - 1546. [Full Text] [PDF] |
||||
![]() |
D. H. Adams, A. C. Anyanwu, J. Chikwe, and F. Filsoufi The Year in Cardiovascular Surgery J. Am. Coll. Cardiol., June 23, 2009; 53(25): 2389 - 2403. [Full Text] [PDF] |
||||
![]() |
S. R. Dixon, C. L. Grines, and W. W. O'Neill The year in interventional cardiology. J. Am. Coll. Cardiol., June 2, 2009; 53(22): 2080 - 2097. [Full Text] [PDF] |
||||
![]() |
H. C. Herrmann Transcatheter Aortic Valve Implantation: Past, Present, and Future Circ Cardiovasc Interv, December 1, 2008; 1(3): 159 - 160. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|