Defining Prolonged Dwell Time: When Are Advanced Inferior Vena Cava Filter Retrieval Techniques Necessary?

An Analysis in 762 Procedures
Originally publishedhttps://doi.org/10.1161/CIRCINTERVENTIONS.116.003957Circulation: Cardiovascular Interventions. 2017;10:e003957

    Abstract

    Background—

    Despite growth in placement of retrievable inferior vena cava filters, retrieval rates remain low. Filters with extended implantation times present a challenge to retrieval, where standard techniques often fail. The development of advanced retrieval techniques has positively impacted retrieval of retrievable inferior vena cava filters with prolonged dwell times; however, there is no precise definition of the time point when advanced techniques become necessary. We aim to define prolonged retrievable inferior vena cava filters dwell time by determining the inflection point when the risk of standard retrieval technique failure increases significantly, necessitating advanced retrieval techniques to maintain overall technical success of retrieval.

    Methods and Results—

    From January 2009 to April 2015, 762 retrieval procedures were identified from a prospectively acquired database. We assessed patient age/sex, filter dwell time, procedural technical success, the use of advanced techniques, and procedure-related adverse events. Overall retrieval success rate was 98% (n=745). When standard retrieval techniques failed, advanced techniques were used; this was necessary 18% of the time (n=138). Logistic regression identified that dwell time was the only risk factor for failure of standard retrieval technique (odds ratio, 1.08; 95% confidence interval, 1.05–1.10; P<0.001). Spline function regression analysis demonstrated that if dwell time exceeded 7 months, the risk of standard technique failure was 40.9%. Adverse events occurred at a rate of 2% (n=18; 15 minor and 3 major).

    Conclusions—

    The necessity of advanced techniques to maintain technical success of retrieval increases with dwell time. Patients with retrievable inferior vena cava filters in place beyond 7 months may benefit from referral to centers with expertise in advanced filter retrieval.

    Introduction

    WHAT IS KNOWN

    • Historically, retrievable inferior vena cava filters with extended implantation times were difficult or impossible to remove.

    • The development of advanced retrieval techniques has greatly improved retrieval rates, particularly of embedded filters with prolonged dwell time.

    What the Study Adds

    • Now that retrieval failure is rare, prior definitions of extended dwell time based on retrieval failure no longer apply.

    • We have found that the likelihood of standard technique failure, and, therefore, the necessity for advanced retrieval techniques to maintain overall retrieval success, increases significantly at 7 months. We have defined this as the inflection point, a potential new definition for extended implantation time.

    • Knowledge of the inflection point may improve overall retrieval rates by prompting referral of patients with retrievable inferior vena cava filters and prolonged dwell times to centers with advanced retrieval expertise.

    Retrievable inferior vena cava filters (rIVCFs) were developed to provide temporary protection from pulmonary embolism while minimizing the risk of filter-related deep venous thrombosis associated with permanent inferior vena cava filters.1,2 In practice, however, the majority of rIVCFs are left in place permanently, with historic retrieval rates as low as 8.5%.35 Although poor clinical follow-up of patients with these devices plays an important role,6 prolonged filter implantation has been associated with retrieval failure rates as high as 43%.711 Furthermore, prolonged rIVCF dwell time is associated with device-related complications, including fracture, migration, organ penetration, and increased risk of deep venous thrombosis.11 These concerns were mirrored in a 2014 safety communication issued by the US Food and Drug Administration urging rIVCF retrieval once no longer indicated.12

    The development of advanced retrieval techniques has had a significant impact on rIVCF retrieval rates, particularly for devices that are in place for extended periods of time or are embedded.1315 However, there is currently no specific definition of what rIVCF dwell time requires increased use of advanced retrieval techniques. We aim to define prolonged rIVCF dwell time by evaluating the time point when the risk of standard retrieval technique failure increases significantly, necessitating advanced retrieval techniques to maintain overall technical success of retrieval. Knowledge of this inflection point could foster removal of prolonged dwell rIVCF via referral to centers with expertise in advanced retrieval techniques, thereby adhering to the Food and Drug Administration mandate and optimizing patient outcomes.

    Methods

    Study Setting and Population

    Our study cohort included all patients referred to our department for rIVCF retrieval from January 2009 to April 2015. The study was institutional review board approved and complied in accordance to the Health Insurance Portability and Accountability Act. All subjects provided written informed consent.

    Data Collection

    Data on patient age/sex, filter dwell time (months), procedural technical success, the use of advanced retrieval techniques, and procedure-related adverse events were collected from a prospectively acquired database. Adverse events were classified as major or minor according to the Society of Interventional Radiology guidelines.16

    rIVCF Retrieval Procedures

    All retrieval procedures were performed by board-certified interventional radiologists, with >90% of procedures performed by 1 of the 3 operators: R.J.L., R.K.R., or K.R.D. Filter retrieval was performed when the rIVCF was no longer indicated for mechanical caval prophylaxis from lower extremity deep venous thrombosis. Technical success is defined as successful endovascular removal of the rIVCF. Standard filter retrieval techniques were initially attempted for all retrieval procedures, consisting of coaxial inner and outer vascular sheaths (Flexor; Cook Medical, Bloomington, IN) with a triloop endovascular snare device (12- to 20-mm EnSnare; Merit Medical Systems, Inc, South Jordan, UT), under conscious sedation per previously published protocols.6 Advanced techniques were used when standard techniques failed to successfully remove the rIVCF. These techniques included directional inner sheath use (7F 55-cm Flexor sheath with Ansel 2 modification; Cook Medical, Bloomington, IN), loop wire snare technique,17,18 balloon disruption/displacement,19 microdissection with endobronchial forceps (Lymol Medical, Woburn, MA),20 and photothermal fibrinolysis using an Excimer laser sheath (Spectranetics Corp, Colorado Springs, CO).21,22

    Statistical Analysis

    Descriptive statistics are reported as means with SD, whereas overall dwell time is reported as a median with 25th and 75th percentiles. Device-specific dwell times are reported as means with SD. Categorical data are expressed as number and percentages. Univariable logistic regression was performed to determine the independent factors associated with failure of standard retrieval technique. Log odds with 95% confidence intervals are reported. Spline logistic regression was used to explore the relationship between dwell time and failure of standard retrieval technique. We first plotted an LOESS curve (locally weighted scatterplot smoothing) of log odds of standard retrieval technique failure versus dwell time; a dwell time of <1 month was used as the reference group. From the LOESS curve, we identified that spline regression was the appropriate regression model. Proper splines (knots) were visually identified. Standard retrieval technique failure was restricted as monotonously increasing with dwell time. Using spline regression, the risk of standard retrieval technique failure was calculated at different time points.23 Statistical significance was determined at the level of 0.05. Statistical analysis was performed with R software (R Core Team, version 3.3.2; Vienna, Austria).

    Results

    From January 2009 to April 2015, 762 retrieval procedures were performed. Mean age of patients in the study cohort was 56 years (SD, 16), and 49% were men. Median rIVCF dwell time across the entire cohort was 1.9 months (25th percentile, 1.2 months; 75th percentile, 3.2 months; range, 0–108.3 months). Filters encountered are listed in Table 1 with mean (SD) dwell times and include the Gunther Tulip, Celect, Celect Platinum (all Cook Medical, Bloomington, IN), Option, Option Elite (both Argon Medical Devices, Inc, Plano, TX), G2, G2X/express, Meridian, Eclipse, Denali (all Bard Peripheral Vascular, Inc, Tempe, AZ), OptEase (Cordis Corp, Miami Lakes, FL), Crux (Volcano Corp, San Diego, CA), and ALN (ALN Implants Chirurgicaux, Ghisonaccia, France).

    Table 1. Filter Types

    FiltersNMean Dwell Time, moSD
    Celect4303.68.3
    Celect Platinum321.6
    Tulip898.219.4
    ALN752.12
    Meridian215.818.7
    Denali152.93.5
    Eclipse119.421.9
    G22817.431.4
    OptEase189.29.4
    Option872.92.6
    Option Elite11.6NA
    Crux34.14.2
    Total7624.611.7

    NA indicates not available.

    Standard retrieval techniques were successful in 82% of procedures (n=626). Advanced retrieval techniques were necessary 18% of the time (n=138). With the use of advanced retrieval techniques, the overall technical success rate increased to 98% (n=745). Univariable logistic regression revealed that dwell time was associated with failure of standard retrieval techniques, thereby increasing necessity of advanced techniques to maintain overall retrieval success (odds ratio, 1.08; 95% confidence interval, 1.05–1.10; P<0.001). Table 2 summarizes the findings from univariable analysis.

    Table 2. Univariable Analysis of Factors Associated With Failure of Standard Retrieval Techniques

    VariablesOdds Ratio (95% CI)P Value
    Age, y0.99 (0.98–1)0.2
    Men0.48 (0.61–1.28)0.5
    Dwell time1.08 (1.05–1.10)<0.001

    CI indicates confidence interval.

    An LOESS scatter plot of the log odds of standard retrieval technique failure versus dwell time was constructed (Figure). From the LOESS curve, we identified 2 splines (dwell times of ≈7 and 30 months). A spline linear function was constructed to describe the relationship between dwell time and the log odds of standard technique failure. The first change in slope was identified at 7 months; the risk of standard technique failure at this point was 40.9%. After 7 months, the risk of standard technique failure remained high and continued to increase, as expected. Characterization of the 30-month spline is limited because of the distribution of the data: in this cohort, the longest dwell time was 108 months. However, >90% of the patients had dwell time <10 months. Second, as shown in the Figure, all patients with dwell time >56 months (black dots) failed standard retrieval technique. Therefore, the right tail of the LOESS curve may not be accurate.

    Figure.

    Figure. Scatter plot (black), locally weighted scatterplot smoothing regression curve (blue), and spline linear regression curve (red) depict the relationship between dwell time and log odds of failure of standard retrieval technique.

    Adverse Events

    A total of 18 procedure-related adverse events (3 major and 15 minor) occurred during the study period, at an overall rate of 2%. Three major events occurred during the study period. One occurred in a patient with a Celect filter in place for 5.9 months. During retrieval, 2 filter struts fractured and remained embedded in the caval wall. These filter struts could not be retrieved and were left in place. The second occurred during retrieval of an OptEase filter with a dwell time of 11.4 months. This device required retrieval from a femoral approach, with subsequent groin hematoma requiring 2 days of inpatient hospitalization for observation but no further treatment. The third occurred in a patient with a Celect filter in place for 0.5 months. During retrieval, significant force, along with dual jugular and femoral venous access, was necessary to collapse and remove the filter, resulting in injury to the inferior vena cava with extravasation. This resolved with a short period of balloon tamponade. However, the patient required 3 additional days of inpatient hospitalization because of a groin hematoma. The remaining adverse events were minor, with most being venous access site issues that required nominal or no therapy.

    Discussion

    In the present study, we have found that after a 7-month rIVCF dwell time, the likelihood of standard retrieval technique failure is high, with a calculated risk of 40.9%, and continues to increase with dwell time. Therefore, after 7 months, advanced filter retrieval techniques are necessary to maintain high overall rates of retrieval technical success.

    Given that a large proportion of rIVCFs remain in situ indefinitely,3 it is particularly troubling that there have been increasing reports of complications associated with these devices, including filter fracture, migration, organ penetration, and increased risk of deep venous thrombosis.11,24 In a review of the Food and Drug Administration’s own Manufacturer and User Facility Device Experience database, Andreoli et al25 reported that ≈87% of user-reported adverse events were associated with rIVCFs, as opposed to ≈13% associated with permanent inferior vena cava filters. Similarly, a retrospective review by Desai et al26 found a complication rate of 9% in a series of 449 rIVCFs. Furthermore, studies have demonstrated that device-related rIVCF complications are positively associated with dwell time.2729 Devices that have been in place for extended periods of time potentially encounter prolonged exposure to multivector respiration-based caval biomechanics, resulting in metal fatigue, thereby increasing the risk of fracture and en bloc or component migration/embolization.30 Thus, because of low retrieval rates, any potential benefit of a device optimally designed for limited use may be offset by the increased risk of complications associated with prolonged or often indefinite implantation.

    Prolonged rIVCF implantation poses a significant challenge for retrieval. Several studies have identified prolonged filter dwell time as an independent negative predictor for successful rIVCF retrieval. In a series of 200 patients using 3 different rIVCF devices, Geisbusch et al10 described a dwell time of >90 days as being associated with retrieval failure, similar to results reported by Averginos et al31 and Marquess et al.9 Glocker et al7 found that Cook Celect and Tulip filters with dwell times of >117 days correlated with rIVCF retrieval failure. Of note, these studies except for Marquess et al reported the use of various advanced retrieval techniques when standard techniques were insufficient; however, these techniques were limited to balloon disruption, use of multiple endovascular snares, snare-over-guidewire techniques (similar to loop wire technique), or filter displacement with a reverse curve catheter. Reported successful retrieval rates from these studies ranged from 76% to 92%. In the present study, similar techniques along with the use of rigid endobronchial forceps and Excimer laser sheath–assisted photothermal ablation allowed an overall 98% retrieval success rate in a large cohort.

    The development of advanced retrieval techniques has had a significant impact on overall rIVCF retrieval rates. In a series of 50 patients, Kuo et al15 demonstrated that complex retrieval methods could retrieve filters with 100% success where conventional methods had previously failed. Recent studies have demonstrated that the ability to successfully remove rIVCF is independent of filter dwell time and is positively impacted by the use of advanced retrieval techniques when necessary.13 However, attempts at quantifying the time period that constitutes prolonged dwell time have thus far been largely device specific,32 making generalizations about optimal retrieval windows difficult. Furthermore, previous descriptions of prolonged dwell time were limited to retrieval failure. In the present study, we have defined prolonged dwell time as the time when standard retrieval techniques fail at high rates; thereby requiring advanced retrieval techniques to maintain overall retrieval success. At 7 months, while encountering a variety of rIVCF, the calculated risk of standard technique failure (and, therefore, requirement of advanced retrieval techniques) is 40.9%. This relationship continues to increase with dwell time. Filters with prolonged dwell times often require complex procedures and the use of multiple advanced techniques. Thus, attempts at retrieval of these devices at centers with limited experience in these techniques may be predisposed to fail.

    Previously published studies have reported higher procedure-related adverse event rates when using advanced techniques. Al-Hakim et al14 reported a complication rate of 5.3% when using advanced retrieval techniques. In the present study, our overall complication rate was 2%; in a previous analysis, we found no association between adverse event rates and filter dwell time or use of advanced techniques on multivariable analysis.13 With clinical experience, it seems that rIVCF retrieval procedures have an excellent safety profile.

    There are limitations to this study. The use of specific advanced techniques was based on operator preference and was not objectively assigned. Another limitation is the single-center nature of our experience, thereby limiting broader clinical application. However, as demonstrated in our analysis and previously published studies, retrieval of rIVCF with prolonged dwell time often requires complex techniques and optimally should be performed in a center with a high level of experience in these procedures. The distribution of the data set is not normal; >90% of the cohort had an rIVCF in place for ≤10 months. Therefore, characterization of the second spline at a 30-month dwell time is limited. Further study is necessary to characterize dwell time in patients with prolonged dwell times (eg, >10 months). Finally, multiple types of rIVCF were encountered in the study period. The technical challenges encountered during a retrieval procedure vary by device33; as a result, the data may vary by device type. In the current cohort, comparisons of the impact of device subtype on the inflection point would be potentially confounded because of the disparate number of filters encountered. As further data are gathered, future studies could be performed to examine the inflection point for necessity of advanced retrieval techniques by device subtype.

    Conclusions

    Retrieval of rIVCF with prolonged dwell time is historically unsuccessful because of lack of expertise in advanced retrieval techniques. Based on our data, we have defined prolonged dwell time as 7 months, where there is significant risk of standard technique failure, thereby requiring advanced techniques to maintain a high rate of technical success. Knowledge of this inflection point may be particularly useful in guiding referral of patients to centers with expertise in advanced retrieval techniques, where rIVCF with extended implantation times can be safely removed. As data on the potential risks of prolonged dwell rIVCFs continue to mount, it is important to optimize opportunities to remove rIVCFs that are no longer indicated.

    Footnotes

    Correspondence to Robert J. Lewandowski, MD, Department of Radiology, Northwestern University Feinberg School of Medicine, 676 N St Clair St, Suite 800, Chicago, IL 60611. E-mail

    References

    • 1. Decousus H, Leizorovicz A, Parent F, Page Y, Tardy B, Girard P, Laporte S, Faivre R, Charbonnier B, Barral FG, Huet Y, Simonneau G. A clinical trial of vena caval filters in the prevention of pulmonary embolism in patients with proximal deep-vein thrombosis. Prevention du Risque d’Embolie Pulmonaire par Interruption Cave Study Group.N Engl J Med. 1998; 338:409–415. doi: 10.1056/NEJM199802123380701.CrossrefMedlineGoogle Scholar
    • 2. Group PS. Eight-year follow-up of patients with permanent vena cava filters in the prevention of pulmonary embolism: the PREPIC (Prevention du Risque d’Embolie Pulmonaire par Interruption Cave) randomized study.Circulation. 2005; 112:416–422.LinkGoogle Scholar
    • 3. Sarosiek S, Crowther M, Sloan JM. Indications, complications, and management of inferior vena cava filters: the experience in 952 patients at an academic hospital with a level I trauma center.JAMA Intern Med. 2013; 173:513–517. doi: 10.1001/jamainternmed.2013.343.CrossrefMedlineGoogle Scholar
    • 4. Helling TS, Kaswan S, Miller SL, Tretter JF. Practice patterns in the use of retrievable inferior vena cava filters in a trauma population: a single-center experience.J Trauma. 2009; 67:1293–1296. doi: 10.1097/TA.0b013e3181b0637a.CrossrefMedlineGoogle Scholar
    • 5. Ray CE, Mitchell E, Zipser S, Kao EY, Brown CF, Moneta GL. Outcomes with retrievable inferior vena cava filters: a multicenter study.J Vasc Interv Radiol. 2006; 17:1595–1604. doi: 10.1097/01.RVI.0000239102.02956.65.CrossrefMedlineGoogle Scholar
    • 6. Minocha J, Idakoji I, Riaz A, Karp J, Gupta R, Chrisman HB, Salem R, Ryu RK, Lewandowski RJ. Improving inferior vena cava filter retrieval rates: impact of a dedicated inferior vena cava filter clinic.J Vasc Interv Radiol. 2010; 21:1847–1851. doi: 10.1016/j.jvir.2010.09.003.CrossrefMedlineGoogle Scholar
    • 7. Glocker RJ, Novak Z, Matthews TC, Patterson MA, Jordan WD, Pearce BJ, Passman MA. Factors affecting Cook Gunther Tulip and Cook Celect inferior vena cava filter retrieval success.J Vasc Surg Venous Lymphat Disord. 2014; 2:21–25. doi: 10.1016/j.jvsv.2013.09.002.CrossrefMedlineGoogle Scholar
    • 8. Dinglasan LA, Oh JC, Schmitt JE, Trerotola SO, Shlansky-Goldberg RD, Stavropoulos SW. Complicated inferior vena cava filter retrievals: associated factors identified at preretrieval CT.Radiology. 2013; 266:347–354. doi: 10.1148/radiol.12120372.CrossrefMedlineGoogle Scholar
    • 9. Marquess JS, Burke CT, Beecham AH, Dixon RG, Stavas JM, Sag AA, Koch GG, Mauro MA. Factors associated with failed retrieval of the Gunther Tulip inferior vena cava filter.J Vasc Interv Radiol.2008; 19:1321–1327. doi: 10.1016/j.jvir.2008.06.004.CrossrefMedlineGoogle Scholar
    • 10. Geisbusch P, Benenati JF, Pena CS, Couvillon J, Powell A, Gandhi R, Samuels S, Uthoff H. Retrievable inferior vena cava filters: factors that affect retrieval success.Cardiovasc Intervent Radiol. 2012; 35:1059–1065. doi: 10.1007/s00270-011-0268-x.CrossrefMedlineGoogle Scholar
    • 11. McLoney ED, Krishnasamy VP, Castle JC, Yang X,, Guy G. Complications of Celect, Gunther tulip, and Greenfield inferior vena cava filters on CT follow-up: a single-institution experience.J Vasc Interv Radiol. 2013; 24:1723–1729.CrossrefMedlineGoogle Scholar
    • 12. Administration USFaD. Removing Retrievable Inferior Vena Cava Filters: FDA Safety Communication.https://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm396377.htm. Accessed March 9, 2016.Google Scholar
    • 13. Desai KR, Lewandowski RJ, Salem R, Mouli SK, Karp JK, Laws JL, Ryu RK. Retrieval of inferior vena cava filters with prolonged dwell time: a single-center experience in 648 retrieval procedures.JAMA Intern Med. 2015; 175:1572–1574. doi: 10.1001/jamainternmed.2015.2561.CrossrefMedlineGoogle Scholar
    • 14. Al-Hakim R, Kee ST, Olinger K, Lee EW, Moriarty JM, McWilliams JP. Inferior vena cava filter retrieval: effectiveness and complications of routine and advanced techniques.J Vasc Interv Radiol. 2014; 25:933–939; quiz 940. doi: 10.1016/j.jvir.2014.01.019.CrossrefMedlineGoogle Scholar
    • 15. Kuo WT, Robertson SW, Odegaard JI, Hofmann LV. Complex retrieval of fractured, embedded, and penetrating inferior vena cava filters: a prospective study with histologic and electron microscopic analysis.J Vasc Interv Radiol. 2013; 24:622–630.e1; quiz 631.CrossrefMedlineGoogle Scholar
    • 16. Sacks D, McClenny TE, Cardella JF, Lewis CA. Society of interventional radiology clinical practice guidelines.J Vasc Interv Radiol. 2003; 14(9 pt 2):S199–S202.CrossrefMedlineGoogle Scholar
    • 17. Esparaz AM, Ryu RK, Gupta R, Resnick SA, Salem R, Lewandowski RJ. Fibrin cap disruption: an adjunctive technique for inferior vena cava filter retrieval.J Vasc Interv Radiol. 2012; 23:1233–1235. doi: 10.1016/j.jvir.2012.06.001.CrossrefMedlineGoogle Scholar
    • 18. Rubenstein L, Chun AK, Chew M, Binkert CA. Loop-snare technique for difficult inferior vena cava filter retrievals.J Vasc Interv Radiol. 2007; 18:1315–1318. doi: 10.1016/j.jvir.2007.07.002.CrossrefMedlineGoogle Scholar
    • 19. Lynch FC. Balloon-assisted removal of tilted inferior vena cava filters with embedded tips.J Vasc Interv Radiol. 2009; 20:1210–1214. doi: 10.1016/j.jvir.2009.06.022.CrossrefMedlineGoogle Scholar
    • 20. Stavropoulos SW, Dixon RG, Burke CT, Stavas JM, Shah A, Shlansky-Goldberg RD, Trerotola SO. Embedded inferior vena cava filter removal: use of endobronchial forceps.J Vasc Interv Radiol. 2008; 19:1297–1301. doi: 10.1016/j.jvir.2008.04.012.CrossrefMedlineGoogle Scholar
    • 21. Kuo WT, Odegaard JI, Rosenberg JK, Hofmann LV. Excimer laser-assisted removal of embedded inferior vena cava filters: a single-center prospective study.Circ Cardiovasc Interv. 2013; 6:560–566. doi: 10.1161/CIRCINTERVENTIONS.113.000665.LinkGoogle Scholar
    • 22. Kuo WT, Cupp JS. The excimer laser sheath technique for embedded inferior vena cava filter removal.J Vasc Interv Radiol. 2010; 21:1896–1899. doi: 10.1016/j.jvir.2010.08.013.CrossrefMedlineGoogle Scholar
    • 23. Lv J, Yang Y, Zhang H, Chen W, Pan X, Guo Z, Wang C, Li S, Zhang J, Zhang J, Liu L, Shi S, Wang S, Chen M, Cui Z, Chen N, Yu X, Zhao M, Wang H. Prediction of outcomes in crescentic IgA nephropathy in a multicenter cohort study.J Am Soc Nephrol. 2013; 24:2118–2125. doi: 10.1681/ASN.2012101017.CrossrefMedlineGoogle Scholar
    • 24. Jia Z, Wu A, Tam M, Spain J, McKinney JM, Wang W. Caval penetration by inferior vena cava filters: a systematic literature review of clinical significance and management.Circulation. 2015; 132:944–952. doi: 10.1161/CIRCULATIONAHA.115.016468.LinkGoogle Scholar
    • 25. Andreoli JM, Lewandowski RJ, Vogelzang RL, Ryu RK. Comparison of complication rates associated with permanent and retrievable inferior vena cava filters: a review of the MAUDE database.J Vasc Interv Radiol. 2014; 25:1181–1185. doi: 10.1016/j.jvir.2014.04.016.CrossrefMedlineGoogle Scholar
    • 26. Desai TR, Morcos OC, Lind BB, Schindler N, Caprini JA, Hahn D, Warner D, Gupta N. Complications of indwelling retrievable versus permanent inferior vena cava filters.J Vasc Surg Venous Lymphat Disord. 2014; 2:166–173. doi: 10.1016/j.jvsv.2013.10.050.CrossrefMedlineGoogle Scholar
    • 27. Vijay K, Hughes JA, Burdette AS, Scorza LB, Singh H, Waybill PN, Lynch FC. Fractured bard recovery, G2, and G2 express inferior vena cava filters: incidence, clinical consequences, and outcomes of removal attempts.J Vasc Interv Radiol. 2012; 23:188–194. doi: 10.1016/j.jvir.2011.10.005.CrossrefMedlineGoogle Scholar
    • 28. Nicholson W, Nicholson WJ, Tolerico P, Taylor B, Solomon S, Schryver T, McCullum K, Goldberg H, Mills J, Schuler B, Shears L, Siddoway L, Agarwal N, Tuohy C. Prevalence of fracture and fragment embolization of Bard retrievable vena cava filters and clinical implications including cardiac perforation and tamponade.Arch Intern Med. 2010; 170:1827–1831. doi: 10.1001/archinternmed.2010.316.CrossrefMedlineGoogle Scholar
    • 29. Angel LF, Tapson V, Galgon RE, Restrepo MI, Kaufman J. Systematic review of the use of retrievable inferior vena cava filters.J Vasc Interv Radiol. 2011; 22:1522–1530.e3. doi: 10.1016/j.jvir.2011.08.024.CrossrefMedlineGoogle Scholar
    • 30. Kuo WT, Robertson SW. Bard denali inferior vena cava filter fracture and embolization resulting in cardiac tamponade: a device failure analysis.J Vasc Interv Radiol. 2015; 26:111–115.e1. doi: 10.1016/j.jvir.2014.08.001.CrossrefMedlineGoogle Scholar
    • 31. Avgerinos ED, Bath J, Stevens J, McDaniel B, Marone L, Dillavou E, Cho JS, Makaroun MS,, Chaer RA. Technical and patient-related characteristics associated with challenging retrieval of inferior vena cava filters.Eur J Vasc Endovasc Surg. 2013; 46:353–359.CrossrefMedlineGoogle Scholar
    • 32. Imberti D, Ageno W, Manfredini R, Fabbian F, Salmi R, Duce R, Gallerani M. Interventional treatment of venous thromboembolism: a review.Thromb Res. 2012; 129:418–425. doi: 10.1016/j.thromres.2011.11.003.CrossrefMedlineGoogle Scholar
    • 33. Chick JF, Stavropoulos SW, Shin BJ, Shlansky-Goldberg RD, Mondschein JI, Sudheendra D, Nadolski GJ, Watts MM, Trerotola SO. A 16-F sheath with endobronchial forceps improves reported retrieval success of long-dwelling “closed cell” inferior vena cava filter designs.J Vasc Interv Radiol. 2016; 27:1027–1033. doi: 10.1016/j.jvir.2016.03.047.CrossrefMedlineGoogle Scholar